Yes, you can load the sample onto the µPAC™ column at a higher flow rate. For example, when you perform your gradient separation at 300 nL/min, you can load the sample at 750 nL/min. As a consequence of the low backpressure of our columns, this is feasible as you will not reach the upper pressure limit (350 bar or 5,000 psi).